• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US STD NECK SEGMENT CORAIL AMT; FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US STD NECK SEGMENT CORAIL AMT; FEMORAL TRIALS Back to Search Results
Catalog Number L20431
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.Investigation summary : it was reported that the neck segment is damaged.No patient implications or harm.The product was returned to depuy synthes for evaluation.Visual inspection revealed scratches all over the std neck segment corail amt surface, blocking device information.Additionally, some deformations were observed on the head of device.This type of damage is consistent with other tools and hard edges coming in contact with the device.Deformations can be traced to improper handling or care of device.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the std neck segment corail amt would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to an unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing records evaluation (mre) was not performed as no lot number was able to be retrieved for this device.If the lot/serial number becomes available, the record will be re-assessed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the neck segment is damaged.No patient implications or harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STD NECK SEGMENT CORAIL AMT
Common Device Name
FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19810627
Report Number1818910-2024-15688
Device Sequence Number15540908
Product Code LXH
UDI-Device Identifier10603295258216
UDI-Public(01)10603295258216
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberL20431
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2024
Initial Report FDA Received Date07/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
S-ROM*ADAPTOR HUDSON TO ZIMMER
Patient SexUnknown
-
-