• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 8.0MM/3.2MM DRILL SLEEVE 200MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 8.0MM/3.2MM DRILL SLEEVE 200MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 03.010.064
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.D4: udi: as the lot and serial number for the device involved in the event was not provided, the full udi is currently not available.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j sales representative.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.
 
Event or Problem Description
It was reported that on june 28, 2024, the patient underwent orif surgery for fracture of the proximal humerus on proximal humerus with the products in question.In the surgery, during the drilling of the distal lateral stop, the drill bits in question were not sharp enough, although the patient's bone quality was also a factor.The sleeve in question was inserted and drilled until it contacted the bone, avoiding the soft tissue, but the drill interfered in the g-hole.The image confirmed that the drill tip reached inside the nail hole, so when the drill tip was lightly tapped with a hammer, the drill tip passed through the hole.Finally, when inserting the end cap in question, the end cap was reinserted several times, but it did not engage properly, and the surgeon decided to close the wound with the cap engaged at an angle.The surgery was completed successfully with a 30 minute surgical delay.No further information is available at this time.This report is for one (1) 8.0mm/3.2mm drill sleeve 200mm.This is report 5 of 6 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8.0MM/3.2MM DRILL SLEEVE 200MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19810758
Report Number8030965-2024-09132
Device Sequence Number18229914
Product Code LXH
UDI-Device Identifier07611819760912
UDI-Public(01)07611819760912
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.064
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/27/2024
Initial Report FDA Received Date07/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3.2 3-FLTD DRILL BIT QC/330/100 CALIB; DRILL BIT Ø3.2 CALIBR L340 3FLUTE F/QUIC; MULTILOC END CAP F/MULTILOC HN/PHN EXTEN; UNK - GUIDES/SLEEVES/AIMING: SLEEVE
Patient Age55 YR
Patient SexFemale
Patient Weight84 KG
-
-