| Model Number |
V+O+C+S+N, JAPANESE |
| Medical Device Problem Codes |
Use of Device Problem (1670); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: the authorized third party service provider (asp) will further evaluated and repair the device.A follow-up report will be submitted when the investigation is complete as defined by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
An authorized third party service provider (asp) contacted ventec to report that the device had low inspiratory pressure.There were no reports of patient involvement associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: ventec has determined that this medwatch report was submitted in error.The reported issue of low inspiratory pressure was incorrect.The initial reporter, an authorized service provider (asp), was performing preventative maintenance on the device and failed to allow it adequate time to warm up prior to testing.Once the device warmed up, proper device operation was confirmed.As a result, this does not meet reportability requirements as the issue would not cause or contribute to death or serious injury if it were to recur.
|
| |
|