MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number RLT311413
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abdominal Pain (1685); Chest Pain (1776); Vascular Dissection (3160)
Date of Event 06/20/2024
Type of Reportable Event Death
Additional Manufacturer Narrative
H3: other code and h6: code b20: the device remains implanted and therefore no product evaluation has been performed.Images was not provided.As part of our routine quality procedures, each batch of devices undergoes comprehensive quality control testing and inspections prior to release for distribution.Gore reviewed the manufacturing records associated with the reported lot number and verified that all release criteria had been met.Based on the incident description and the subsequent investigation, no further information was provided to gore, we are unable to determine the cause of this incident and assign a root cause.Per gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention or additional intraoperative procedure time include, but are not limited to: dissection, perforation, or rupture of the aortic vessel and surrounding vasculature.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
It was reported to gore that a gore® excluder® aaa endoprosthesis (rlt311413) and a gore® excluder® iliac branch endoprosthesis (ceb231410) was implanted on (b)(6) 2024.Everything went wonderfully and the patient went to the recovery room without any abnormalities.A few days later the patient complained of abdominal pain.The ct of the abdomen was normal.Another few days later the patient suffered severe pain in the chest area and the ct showed a type b dissection that went all the way down to the common iliac artery and compressed the rlt311413.The patient was given an emergency thoracic endovascular aortic repair (tevar) with a non-gore device (medtronoic device 42/157 mm), then transferred to intensive care, where the patient passed away a few days later.No device explant has been performed.It is reported that the patient had a hypertensive crisis, which is considered the main reason for a type b dissection.
 
Additional Manufacturer Narrative
Several attempts have been made to obtain imaging for evaluation, however, no images were not provided to gore.Due to the hypertensive crisis, which is considered the main reason for a type b dissection, therefore it is reasonable to believe that a patient condition may have contributed to the dissection which lead to the death of the patient.Remove medical device problem code a24 as it was inadvertently entered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
ida simson
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key19812308
Report Number3007284313-2024-03382
Device Sequence Number8380211
Product Code MIH
UDI-Device Identifier00733132619078
UDI-Public00733132619078
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberRLT311413
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2024
Supplement Date Received by Manufacturer09/27/2024
Initial Report FDA Received Date07/23/2024
Supplement Report FDA Received Date09/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS
Outcome Attributed to Adverse Event Required Intervention; Death;
Patient Age67 YR
Patient SexMale
Patient Weight135 KG
-
-