• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 5.0MM, DRL BIT, CANNULATED, REUSABLE; DRILL BIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 5.0MM, DRL BIT, CANNULATED, REUSABLE; DRILL BIT Back to Search Results
Model Number 5.0MM, DRL BIT, CANNULATED, REUSABLE
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
It was reported that during a sub-talar joint fusion there were two blunt 5mm drills (ar-8770-02-ru) on the 5/7 mm comp ft set.There was no harm for patient, operator or third party.The surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Additional information: based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to wear and tear incurred from repeated usage/reprocessing.Manufactured date: 20-jun-2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0MM, DRL BIT, CANNULATED, REUSABLE
Common Device Name
DRILL BIT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19815258
Report Number1220246-2024-06865
Device Sequence Number8695170
Product Code HTW
UDI-Device Identifier00888867319592
UDI-Public00888867319592
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5.0MM, DRL BIT, CANNULATED, REUSABLE
Device Catalogue NumberAR-8770-02-RU
Device Lot Number032224
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2024
Supplement Date Received by Manufacturer06/06/2025
Initial Report FDA Received Date07/24/2024
Supplement Report FDA Received Date07/01/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/20/2022
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-