| Model Number |
10762472 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3, h6: initial corrective actions/preventive actions implemented by the manufacturer: the system was taken out of use as instructed by customer service advisory notification (csan) xp017/24/s which began distribution to affected customers on 07/02/2024.Siemens initiated recall res# 94948, which was submitted to the fda on 07/05/2024 for potential problems with the support arm of select system ceiling displays or wall displays.Manufacturers preliminary analysis: the root cause for the issue has not yet been determined.The investigation is ongoing.The root cause will be communicated via res# 94948.Corrective and or preventative actions will be implemented via resolution updates xp018/24/s and, if necessary, xp019/24/s.These updates are anticipated to be released in september 2024.No supplemental report for this complaint will be submitted to the fda.
|
| |
|
Event or Problem Description
|
|
Siemens healthineers was notified of an issue with the luminos agile max system.It was communicated that the customer visually inspected the mounting of the support arm of the ceiling suspension per siemens healthineers customer safety advisory notification xp017/24/s.The customer identified that the mounting screw in the support arm of the display ceiling suspension was missing.The monitor support arm remained attached to the ceiling and no injury occurred due to this issue.It is assumed that in a worst-case scenario, a serious injury could result if a person was located beneath the display monitor during adjustment, and it suddenly lowered.Injury could be sustained from the falling equipment.
|
| |
|
Search Alerts/Recalls
|