| Model Number |
PS210-030S-LIGHT |
| Medical Device Problem Codes |
Fire (1245); Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a generator.It was reported that the handpiece caught on fire during a surgery.The site noticed smoke and small flames coming from the handpiece.There was a delay of less than 1 hour and no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: no parts have been received by the manufacturer for evaluation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis summary: complaint unconfirmed.Upon receiving the device, it looked to be unused with no damage.The device was decontaminated per wi and then tested per wi.Upon testing, the handpiece did work correctly and there was not any smoke or small flames that formed.The coating is worn on the blade.H6: codes b01, c19 <(>&<)> d14 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|