| Catalog Number |
0620040610 |
| Medical Device Problem Codes |
Increase in Pressure (1491); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was overpressure during procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that there was overpressure during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: actual pressure on the unit was reading 0 and the insufflator kept insufflating.Probable root cause: 1.Pressure sensor malfunction / out of calibration.2.Software malfunction.3.Use error.4.System design.5.Unwanted movement of internal components / wiring.6.Pressure button does not disengage.7.Electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge.8.Venting valve system or overpressure alarm failure.9.Safety valve or regulator malfunction.10.Hpu or lpu assembly malfunction.11.Ppv failure.12.Manufacturing/ service error.The reported failure mode will be monitored for future reoccurrence.The manufacture date is not known.
|
| |
|