| Catalog Number |
81324 |
| Medical Device Problem Codes |
Restricted Flow rate (1248); No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
While using a cavitron select sps g124, they allege that they have no water , the solenoid is clogged and the handpiece was getting hot no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Rb 8/06/2024.Hpc lot # - 01230 04240.St lot # - 202303.The hpc is damaged, restricting water flow to the st, debris in water filter, exposed wires on hpc.Damaged parts have been replaced.Unit have been calibrated and tested to factory's specs.No other faults found.
|
| |
|