VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VOCSN, ENGLISH
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was evaluated by ventec where the reported issues of it measuring lower peep (positive end expiratory pressure) than set and having low exhaled tidal volume (vte) levels was confirmed.Ventec replaced the internal flow transducer (ift) to resolve the reported issues.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issues was the ift.
 
Event or Problem Description
It was reported that the device needed service.During the device evaluation ventec observed that the ventilator measured lower peep (positive end expiratory pressure) than set and that its exhaled tidal volume (vte) levels were low.There was no patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer Contact
elizabeth gilbert
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key19819199
Report Number3013095415-2024-00538
Device Sequence Number8671190
Product Code CBK
UDI-Device Identifier00855573007006
UDI-Public(01)00855573007006(11)180130
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K162877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberVOCSN, ENGLISH
Device Catalogue NumberPRT-00490-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2024
Initial Report FDA Received Date07/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/30/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-