| Model Number |
9735542 |
| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H2: correction - updated h5 and h8 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a thermal therapy system being used during a soft tissue ablation (neuro) procedure.It was reported that they experienced a noisier than usual temperature map.The mri had produced noisier image than usual that same day in a previously logged case using flex coils, but in this case we used the standard head coil.Similar imaging artifacts were observed in the map, but it was worse in some areas.It was later determined that their head ring they left the patient in had a component that caused additional imaging artifacts.The area of ablation was not impacted by artifacts, but the imaging just looked noisy and less representative of their usual experience.Ablations proceeded successfully.Upon removal of the first of two catheters it was noted the tip was melted and post-imaging showed a little blood in the same region where the tip would have melted which is consistent with our understandings.No expressed patient impact as of close of procedure and no delays experienced.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: section d information references the main component of the system.Other relevant device(s) are: product id: m450001b015, version: 3.2 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: correction - mdr 1723170-2024-01786 is against the vclas device 9735560 and therefore is related to the vclas melt/bleed and not the system/software related issue (¿noise on the tmap¿).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information in b5.Correction to b3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received.It was reported that the site had since determined that many of their relative tmap imaging artifact issues were due to the anesthesia machine being too close to the gauss line.Subsequent cases have been better.
|
| |
|
Additional Manufacturer Narrative
|
|
A software investigation analysis was initiated to determine the probable cause of the issue through log analysis.Analysis found that the reported issue was not cause by the software.The head ring left on the patient caused the perceived image noisiness in the sessions.Two sessions involved a high laser power applications that lasted longer than recommended.Analysis found that the software functioned as designed.Codes b01, c19, d14 are applicable to this analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|