|
The serial number of the cobas pure c 303 analytical unit is (b)(6).On the field service representative's (fsr) first service visit, he changed the cuvettes and the lamps.On the fsr's second service visit, he reviewed the calibrations and qc data; no issues were found.The software was also updated.On the fsr's third service visit (together with the field applications specialist fsa) the analyzer's maintenance was checked.The fsr performed blank and photometric checks with acceptable results.They also replaced the reagent pack and performed a calibration with satisfactory results.A 20-repetition precision check using patient samples was performed with only one sample having an unacceptable result.The fsr and fsa verified that the analyzer was performing within specifications.On the fsr's fourth service visit, he decontaminated the incubation bath with chlorine, changed the incubation water six times to eliminate dirt, replaced the heat filter, cleaned the washing comb pipes, and verified the cuvette washes.He also performed photometers and cuvette blanks with acceptable results.The fsa performed checks and ran 90 routine samples for correlation with satisfactory results.On the fsr's fifth service visit, he disinfected the incubation bath, cleaned the washing comb, reviewed the cuvette washing, changed the heat filter, and cleaned the sample probe.Operational checks and 90 patient sample runs were again performed with satisfactory results.The investigation is ongoing.
|
|
The initial reporter received questionable triglycerides results from seven patient samples tested on the cobas pure c 303 analytical unit.The initial results were not reported outside of the laboratory as they did not match the patients' medical histories.The reporter deemed all results incorrect.On (b)(6) 2024: sample (b)(6).The initial result was 582 mg/dl.The repeat result was 374 mg/dl.On (b)(6) 2024: sample (b)(6).The initial result was 594 mg/dl.The repeat result was 211 mg/dl.On (b)(6) 2024: sample (b)(6).The initial result was 609 mg/dl.The repeat result was 218 mg/dl.On (b)(6) 2024: sample (b)(6).The initial result was 575 mg/dl.The repeat result was 255 mg/dl.On (b)(6) 2024: sample (b)(6).The initial result was 400 mg/dl.The repeat result was 200 mg/dl.On (b)(6) 2024: sample (b)(6).The initial result was 553 mg/dl.The repeat result was 179 mg/dl.The field service representative (fsr) instructed the reporter to change the reagent lot, calibrate, and run qc.The issue was not resolved.On (b)(6) 2024: sample (b)(6).The initial result was 553 mg/dl.The first repeat result was 280 mg/dl.The second repeat result was 181 mg/dl.
|