Section h10: the device was returned for evaluation, device had 5x self-test records.First 3 tests from aug and dec 2023 passed.Tests from 14th and 22nd april failed on pump flow, leak and overvac.Pump current lower than before, but still a clear fail.Pump had 115hrs operation.Device had been giving good therapy until 4th april, then gave continuous leak alerts/alarms, possibly at canister change.Pump cannot reach -125mmhg on a softport clp.Confirmed carbon contamination.There is no indication that this device failed to meet specifications at the time of manufacture.A review of prior escalation actions has revealed that a preliminary risk assessment ¿pra¿ was conducted following complaints for renasys edge failing self-test and noisy operation, the investigation recovered and inspected several devices.The results concluded that black particulate matter from the carbon filter component of the renasys edge canisters has been confirmed to be entering the npwt pump and causing damage.It is suspected the dust filter between the carbon filter and the connection to the pump inlet is failing/ not sufficiently bonded the pra was escalated to health hazard evaluation ¿hhe¿ to assess the risk to the user.The hhe concluded that in the most likely scenario, the canister and pump operate normally.There is no hazardous situation and no harm, and the health risk index is none/negligible.In the worst-case scenario, the canister filter fails during home healthcare use and pump stops functioning.The pump alerts the user to failure or leak alarm.There may be an extended delay in therapy while a replacement pump is sourced.This could potentially lead to wound deterioration or skin graft failure.The health risk index is low.Based on the potential for cumulative effects of the canisters on the pump function, the hhe conclusion is to advance to field action board ¿fab¿.The field action board initiated a product recall on both renasys edge 300 & 800ml canisters with solidifiers.An internal quality hold was also initiated.The root cause of the failure was determined to be that carbon particles from the canister odour filter had passed through the dust filter into the pump unit of the device, resulting in failed self-test, ability to provide suction and can result in excess noise.Non-conformance ¿nc¿ was created to capture the investigation and plan changes with a corrective action currently in place to manage the actions needed to reduce the probability of future occurrences of this nature.Smith and nephew will monitor capa effectiveness and for further adverse trends relating to the reported allegation.S+n is conservatively reporting this event in association with a class ii recall of the concomitant canisters pursuant to applicable regulations.Internal complaint reference number: (b)(4).
|
|
Following a review of the available information, the manufacturer has determined that this event does not meet the criteria for mdr reporting.The device is intended to display an alarm/ alert if it detects an issue and may discontinue applying therapy until the issue is resolved.The loss of negative pressure wound therapy benefits would not be likely to cause or contribute to death or serious injury.Event may cause minor or temporary injury (e.G., delayed wound healing, maceration, etc.) requiring no or minor medical intervention.Event may require use of a backup device to continue therapy.This event did not lead to a death or serious injury, nor would it be likely to cause or contribute to a death or serious injury if it were to recur.This event is considered not reportable pursuant to applicable regulations.
|