• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC CR TIB INSERT SLOPE++, SZ 4, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC CR TIB INSERT SLOPE++, SZ 4, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-49-4009
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Cyst(s) (1800); Pain (1994); Synovitis (2094); Osteolysis (2377); Metal Related Pathology (4530)
Date of Event 01/24/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10.Related: report 2 of 3 and 3 of 3.H11.Additional manufacturer narrative- report 1 of 3.D10.Concomitant devices: 4576645, 02-012-45-4050 - lgc tibial fit tray cem sz 4f / 5t.4590127, 02-010-03-0340 - logic cr femoral cem, right, sz 4.4638500, 200-02-32 - three peg patella 32mm.These devices are used for treatment not diagnosis.There is no additional information available.Pending investigation.
 
Event or Problem Description
As reported via legal documentation the patient had a right total knee arthroplasty and then approximately 7 years later experienced a revision surgery.Revision operative report -preoperative diagnoses: 1.Failed right total knee replacement secondary to polyethylene breakdown and subsequent cystic changes to both the femur and the tibia, resulting in loosening of the right total knee prosthesis.2.Chronic right knee pain as a result of diagnosis #1.The patient was revised to competitor¿s devices.The right knee clinically and radiographically demonstrated failure.Hardware removed: an exactech total knee replacement system was removed.At the time of the removal, there was gross loosening of the femoral and tibial components.There were cystic degenerative changes throughout the tibia and within the femur both medially and laterally, and there was complete breakdown of the posteromedial tibial tray with subsequent metallosis from metal-on-metal contact of the femoral component on the tibial component, resulting in metallosis villonodular synovitis throughout the knee.Encountered villonodular synovitis classically seen with metallosis that was blackened in color.Villonodular synovitic changes in the underside of the bone of the patella where cystic degenerative changes from polyethylene wear debris had been occurring.The patient was awakened and extubated and taken to the recovery room in stable condition.There were no intraoperative complications.There is no other information available.
 
Additional Manufacturer Narrative
1038671-2024-02511, 1038671-2024-02510 this follow-up report is being submitted to relay additional and/or corrected information.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2024-02511, 1038671-2024-02510.The following sections were corrected: b1, b2, d1, d4, h6.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, femoral loosening, tibial loosening and/or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC CR TIB INSERT SLOPE++, SZ 4, 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19822463
Report Number1038671-2024-02509
Device Sequence Number18313927
Product Code JWH
UDI-Device Identifier10885862174512
UDI-Public10885862174512
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2019
Device Catalogue Number02-012-49-4009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2024
Supplement Date Received by Manufacturer12/18/2024
Initial Report FDA Received Date07/24/2024
Supplement Report FDA Received Date12/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/21/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexMale
Patient Weight129 KG
Patient RaceWhite
-
-