| Model Number |
PCF-H180AI |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed during the device inspection, that the colono video scope exhibited foreign objects in air/water tube and cylinder.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation and the information provided, it could not be determined what the foreign material was.There was no damage to the area where the foreign material was detected, and it was unknown if reprocessing was performed according to the instructions for use (ifu).The event can be detected/prevented by following the instructions for use which state: ifu states the detection method in evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 preparation and inspection ifu states the preventative measures in evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories) olympus will continue to monitor field performance for this device.
|
| |
|