| Catalog Number |
AA6316 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: estimated date.
|
| |
|
Event or Problem Description
|
|
According to the available information, during patient indwelling, the balloon burst and fell out.
|
| |
|
Additional Manufacturer Narrative
|
|
Two used samples were received.After desinfection we can observed that these balloons were burst without missing part.Burst issue of silicone balloon is known but according to the available information we cannot conclude on a specific root cause in this case.After receiving this complaint, we searched for other complaints and we found none regarding the lot number 9362253.Quality database was checked and revealed a corrective and preventive action: capa-000152."balloon issues on folysil and silicone prostatic catheters" was opened and monthly monitoring.No corrective action will be implemented as our trend is not increasing and the threshold is not exceeded.The evaluation has showed that risks are adequately controlled and reduced as far as possible in the state of art.Based on this, we can conclude that residual risks are adequately controlled and reduced as far as possible, and the residual risks associated with the use of the product are acceptable when weighed against the benefits to the patient/user.It is concluded that the risks identified are still acceptable and considered as safe.
|
| |
|
Event or Problem Description
|
|
According to the available information, during patient indwelling, the balloon burst and fell out.
|
| |
|
Search Alerts/Recalls
|