Model Number 434887
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2024
Type of Reportable Event Malfunction
Event or Problem Description
A pk papyrus covered stent system was selected for treatment of a type iii perforation in the first obtuse marginal.The stent was implanted on (b)(6) 2024, but was expired on 31-dec-2023.The expired product was chosen, as the situation was life threatening if not used.
 
Additional Manufacturer Narrative
The affected pk papyrus stent system was not returned to biotronik and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form) and the production documentation were reviewed to establish whether a device deficiency contributed to this event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The pk papyrus was used after the use by date which is indicated on the product label and warned against in the ifu.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
 
Additional Manufacturer Narrative
Corrected the initial reporter information.Reference capa# nc-24-004.The affected pk papyrus stent system was not returned to biotronik and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form) and the production documentation were reviewed to establish whether a device deficiency contributed to this event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The pk papyrus was used after the use by date which is indicated on the product label and warned against in the ifu.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK PAPYRUS US 2.5/15
Common Device Name
COVERED CORONARY STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key19822889
Report Number1028232-2024-03906
Device Sequence Number18309275
Product Code NIV
UDI-Device Identifier07640130447158
UDI-Public(01)07640130447158(17)2312
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
H170004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number434887
Device Catalogue Number434887
Device Lot Number01222035
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2024
Supplement Date Received by Manufacturer08/14/2024
11/20/2024
Initial Report FDA Received Date07/24/2024
Supplement Report FDA Received Date08/20/2024
11/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age84 YR
Patient SexFemale
-
-