| Catalog Number |
309620 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Material #: 309620.Batch#: 4080096.It was reported by customer that pharmacy noticed that a certain lot of the bd# 309653 did not have scale markings on them.As their supply chain team began inspecting this product number and lot across the division, they then noticed that bd# 309620 of a certain lot also did not have scale markings.Complaint received via email(s) attached.Pharmacy noticed that a certain lot of the bd# 309653 did not have scale markings on them.As their supply chain team began inspecting this product number and lot across the division, they then noticed that bd# 309620 of a certain lot also did not have scale markings.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Pr (b)(4) follow up: a device history record review was completed by our quality engineer team for provided material number 309620 and lot number 4080096.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.
|
| |
|
Event or Problem Description
|
|
No additional information received.Material #: 309620; batch#: 4080096.It was reported by customer that pharmacy noticed that a certain lot of the bd# 309653 did not have scale markings on them.As their supply chain team began inspecting this product number and lot across the division, they then noticed that bd# 309620 of a certain lot also did not have scale markings.Verbatim: complaint received via email(s) attached.Pharmacy noticed that a certain lot of the bd# 309653 did not have scale markings on them.As their supply chain team began inspecting this product number and lot across the division, they then noticed that bd# 309620 of a certain lot also did not have scale markings.
|
| |
|