| Catalog Number |
CQF7584 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiration date: 12/2026).H11: d2b (dqy;lit).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure, the pta balloon allegedly popped while inflating.Reportedly, the balloon was removed.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one conquest 40 device was returned for evaluation.An in-house presto inflation device was used to inflate the balloon, and water was noted leaking from the balloon.The balloon fibers were stripped and under microscopic examination, two compound ruptures were noted to the balloon.No other functional testing performed.Therefore, the investigation is confirmed for the reported balloon rupture as two compound ruptures were noted to the balloon.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: g3 h11: b5, d2b (dqy;lit), d4 (unique identifier (udi) #), h6 (method, result) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured while inflating.Reportedly, the balloon was removed and the procedure was completed using another device.There was no reported patient injury.
|
| |
|