| Model Number |
25-449-16-91 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
07/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at kls se.During the investigation the product lot number was not identified; therefore, the device history records were not reviewed.The failure root cause cannot be determined due to the device not being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.The reported rpms exceeded the maximum recommended speed found in the ifu.
|
| |
|
Event or Problem Description
|
|
It was reported during initial surgery a twist drill broke during use.A portion of it remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was successfully completed.The results of the investigation have identified no manufacturing issues, no design issues and/or corrective actions necessary at this time.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.The reported rpms exceeded the maximum recommended speed found in the ifu.Lots identified: 33606099 - mfg date 04/03/2024; 33606103 - mfg date 04/08/2024.
|
| |
|