Analysis of the pump found ¿motor gear train anomaly ¿ corrosion and/or wear and/or lubrication¿ and ¿motor gear train anomaly ¿ stall due to shaft/bearing¿.Interrogation of the pump upon receipt indicated that the pump was delivering fentanyl (1,000.0 mcg/ml at 728.8 mcg/day) and bupivacaine (10.0 mg/ml at 7.288 mg/day).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|