Event or Problem Description
|
|
Subject id: (b)(6).Study no: dots.Clinical adverse events received for migration/loosening.Device and procedure(relatedness).Device related: definitely.Procedure related: definitely not.Date of event: 8/jul/2024.Date of implant: information not provided.Date of revision: information not provided.Device location: right.Treatment/impact: information not provided.Geographical location: us.Additional event information: on (b)(6) 2024, medical records note the patient was noted to have had several falls.The patient is experiencing pain, a limp, and swelling.The patient has pseudolaxity, large varus collapse as well.Radiology reports show failure of tibial tray with subsidence.Physical examination shows a significant effusion, significant pseudolaxity, large varus collapse.Femoral component appears to be well fixed as well the patella is tracking centrally, with no tilt or subluxation.The patient has loose tibia with subsidence.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: corrected: h6 (health effect - clinical and health impact code).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|