| Catalog Number |
1105000026 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report summarizes 3 malfunction events, where it was reported the devices experienced fluid leaks onto floor.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.2 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.1 device is pending evaluation.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes 3 malfunction events, where it was reported the devices experienced fluid leaks onto floor.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
1 pending device was not evaluated, but reviewed by data and confirmed the issue.Section h codes have been updated.
|
| |
|
Search Alerts/Recalls
|