| Catalog Number |
9-ASD-026 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a 26 mm amplatzer septal occluder was selected for an implant using a 10f amplatzer trevisio intravascular delivery system.During the procedure, the device gained a cobra shape deformity while getting out from delivery system.Device was removed from the patient prior to released from the delivery cable.The device was return to its normal shape after a few manipulations.The physician tried to use it 2 more times.However, the result was the same couldn't return to its normal shape after manipulations.The patient was refer for surgical atrial septal defect (asd) closure.No interaction with cardiac structures during deployment and no angulation or kink noticed in the delivery system.The patient is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.A returned device assessment could not be performed as the device was not returned for analysis.Possible causes of device deformity include use of the incorrect size delivery system, anatomical interference, and angulation or kink in the delivery system upon deployment.The correct delivery system was used and no anatomical interference, or angulation or kink in the delivery system was reported.Abbott requested for device/issue images to review; this was not provided by the site.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all defined manufacturing specifications.There were no complaints associated with any other devices from the lot.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Search Alerts/Recalls
|