AVANOS MEDICAL INC. MIC GASTROSTOMY FEEDING TUBE WITH ENFIT CONNECTOR; DH EF BALLOON TUBES PRODUCTS Back to Search Results
Model Number 8110-14LV
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/25/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The sample is reported to be available but has not yet been received by the manufacturer.A review of the device history record and udi are in-progress.All information reasonably known as of 23 jul 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
It was reported, "balloon failed, fluid exiting via feed lumen.Tube placed on (b)(6) 2024.Patient complained of pain.Ct scan confirmed tip of tube in stomach, on return tube had fallen out with empty balloon on (b)(6) 2024.Tube retained and renewed internally, fluid leaking into feed lumen.Patient required observation for peritonitis, elongated hospital stay and post-pyloric feeding while at risk of peritonitis.Has now had a further procedure for post pyloric feeding and since then another rig procedure.¿ per additional information received on 09jul2024, ¿no peritonitis, just risk of, plus monitoring and significant pain.¿.
 
Additional Manufacturer Narrative
The product involved in the report has been returned and the investigation remains in progress at this time.All information reasonably known as of 09 sep 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The device history record for lot 30262737 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation and no failures were detected in the device returned.A root cause could not be determined.All information reasonably known as of (b)(6) 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIC GASTROSTOMY FEEDING TUBE WITH ENFIT CONNECTOR
Common Device Name
DH EF BALLOON TUBES PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key19829573
Report Number9611594-2024-00137
Device Sequence Number18023365
Product Code KNT
UDI-Device Identifier00350770445601
UDI-Public(01)00350770445601(240)120001502(11)230530(17)260529(10)30262737
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K842076
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2026
Device Model Number8110-14LV
Device Catalogue NumberN/A
Device Lot Number30262737
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2024
Supplement Date Received by Manufacturer08/23/2024
06/30/2025
Initial Report FDA Received Date07/25/2024
Supplement Report FDA Received Date09/10/2024
07/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age67 YR
Patient SexMale
Patient Weight56 KG
-
-