| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
01/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced refractive surprise.The lens was exchanged with a same length lens but different power and this resolved the problem.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative: refractive surprise, secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: lens was returned in a micro-centrifuge vial, dry.Visual inspection found no visible damage to the lens.Dimensional and functional inspection found the lens within specifications.Claim#: (b)(4).
|
| |
|
Search Alerts/Recalls
|