• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_12.6
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 01/06/2024
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced refractive surprise.The lens was exchanged with a same length lens but different power and this resolved the problem.
 
Additional Manufacturer Narrative
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative: refractive surprise, secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.(b)(4).
 
Additional Manufacturer Narrative
H3 - device evaluation: lens was returned in a micro-centrifuge vial, dry.Visual inspection found no visible damage to the lens.Dimensional and functional inspection found the lens within specifications.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key19833938
Report Number2023826-2024-03359
Device Sequence Number7877973
Product Code QCB
UDI-Device Identifier00840311308910
UDI-Public00840311308910
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model NumberVTICM5_12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2024
Supplement Date Received by Manufacturer10/23/2024
Initial Report FDA Received Date07/25/2024
Supplement Report FDA Received Date10/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age35 YR
Patient SexFemale
-
-