This device is classified as import for export, therefore 510k is not applicable.Model fb-15v is available in the usa with a 510k number k951199.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the biopsy inlet piece loose.Based on the result, we concluded that it was caused due to the excessive force applied on the biopsy inlet piece.In addition, our technician confirmed that the u/d knob broken, the cfb (coherent fiber bundle) broken, the lcb (light carrying bundle) broken, and the insertion flexible tube crushed; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.This defect occurred not during procedure.Based on the technical report ""hr-rpt-0620(control body)"" and/or the risk analysis results, it was evaluated to submit mdr.
|