• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND EPIQ ELITE; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND EPIQ ELITE; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Model Number EPIQ ELITE ULTRASOUND SYSTEM
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2024
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported the swivel mechanism of their epiq elite ultrasound system¿s control panel did not lock properly while transporting the unit within the user facility.The failure occurred outside of clinical use and no patient or user was harmed as a result of the issue.A philips service engineer realigned the bushing to resolve the issue.A thorough engineering investigation has been performed to identify the root cause of the reported issue.The engineering team determined the failure was a hardware issue in the articulating arm latching mechanism preventing the locking solenoid from fully engaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ ELITE
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
thuy nguyen
22100 bothell everett highway
bothell, WA 98021
MDR Report Key19836423
Report Number3019216-2024-00109
Device Sequence Number7816018
Product Code IYN
UDI-Device Identifier00884838099111
UDI-Public00884838099111
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K200304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberEPIQ ELITE ULTRASOUND SYSTEM
Device Catalogue Number795234
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/26/2024
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-