OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE "PLASMABAND", BAND, MEDIUM, 24 FR., 12°/16°, ESG TURIS; ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL
|
Back to Search Results |
|
| Model Number |
WA22721S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the electrode loop handpiece broke off at the distal end.The issue occurred during an unknown therapeutic procedure.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.Based on the results of the investigation, the distal end of the loop was used near or brought into contact with a conductive material, resulting in a rollover.In such a case, a high current flow is concentrated at a single point of the loop wire.The high current has the potential to generate so much electricity for a very short time that the material melts immediately and solidifies again immediately afterwards.Thus, there was no error message from the generator used.The event can be detected/prevented by following the instructions for use (ifu) which state: caution 1: "caution risk of injury to the patient contact between the hf-resection electrode and metal parts, such as other endoscopic equipment, implants or stents can result in sparkover between the hf-resection electrode and the metal part.Always keep a distance of at least 10 mm between the hf-resection electrode and metal parts." caution 2: "caution risk of injury to the patient use extreme caution when using electrosurgery in close proximity to or in direct contact with any metal objects.Do not activate the hf-resection electrode while any portion of the hf-resection electrode tip is in contact with another metal object.This can cause uncontrolled heating of the hf-resection electrode and may result in damage and breakage to the hf-resection electrode tip.If excessive heating or physical forces cause damage to the hf-resection electrode tip, broken device fragments may remain in the body, possibly requiring additional surgery for removal." olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|
|