• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71938-01
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported product has been returned and investigation is pending at this time.A follow-up report will be filed once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
Customer initially reported their abbott diabetes device did not power on.The product was returned and investigated for not being able to power on, however during investigation the device became too hot to handle.This report is being submitted due to the additional issue observed during returned product investigation.There was no report of death, serious injury, or mistreatment associated with this event.Issues involving thermal events occurring with adc freestyle libre and libre 2 readers is associated with report of corrections and removal number 2954323-02/09/23-001-c, submitted to the fda on 09-feb-23.Adc field action fa1005-2023 was issued to the us 09feb2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Common Device Name
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key19839761
Report Number2954323-2024-27234
Device Sequence Number20557835
Product Code PZE
UDI-Device Identifier00357599002008
UDI-Public(01)00357599002008(10)02B026D
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P160030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 07/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number71938-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2023
Initial Report FDA Received Date07/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/17/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
QUALITY RECORD QR869779
Patient Sequence Number1
Patient SexUnknown
-
-