| Catalog Number |
3512-120 |
| Medical Device Problem Code |
Wrong Label (4073)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a pathfinder straight percutaneous rod was found in packaging labelled for a curved rod.This was discovered at the pharmacy before being sent and there was no patient impact.
|
| |
|
Event or Problem Description
|
|
It was reported that a pathfinder straight percutaneous rod was found in packaging labelled for a curved rod.This was discovered at the pharmacy before being sent and there was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the straight rod has been manually bent.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to excessive forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|