• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. PERC STRGHT 5.5 ROD 120MM; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. PERC STRGHT 5.5 ROD 120MM; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 3512-120
Medical Device Problem Code Wrong Label (4073)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event or Problem Description
It was reported that a pathfinder straight percutaneous rod was found in packaging labelled for a curved rod.This was discovered at the pharmacy before being sent and there was no patient impact.
 
Event or Problem Description
It was reported that a pathfinder straight percutaneous rod was found in packaging labelled for a curved rod.This was discovered at the pharmacy before being sent and there was no patient impact.
 
Additional Manufacturer Narrative
Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the straight rod has been manually bent.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to excessive forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERC STRGHT 5.5 ROD 120MM
Common Device Name
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key19841215
Report Number3012447612-2024-00176
Device Sequence Number7878201
Product Code NKB
UDI-Device Identifier00889024335981
UDI-Public(01)00889024335981(10)85MP
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K132884
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3512-120
Device Lot Number85MP
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/26/2024
Supplement Report FDA Received Date01/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/11/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-