| Catalog Number |
650700000000 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report summarizes 4 malfunction events, where it was reported the devices experienced fowler cannot be lowered or false engagement of handle in the upright or horizontal position.There was 1 event with patient involvement; no adverse consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.4 devices are pending evaluation.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Additional Manufacturer Narrative
|
|
2 devices that were pending evaluation were evaluated in the field and the issue was confirmed.The devices were repaired on site and returned to service.1 device that was pending was evaluated in the field but the issue was not confirmed; no defect or malfunction was found.Section h codes have been updated.Because of this, the number of reported events has been changed from 4 to 3.1 device is still pending evaluation.
|
| |
|
Event or Problem Description
|
|
This report summarizes 3 malfunction events, where it was reported the devices experienced fowler cannot be lowered or false engagement of handle in the upright or horizontal position.There was 1 event with patient involvement; no adverse consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device pending evaluation was evaluated in the field and the issue was confirmed.The device was repaired on site and returned to service.Section h codes have been updated to reflect this.
|
| |
|
Event or Problem Description
|
|
This report summarizes 3 malfunction events, where it was reported the devices experienced fowler cannot be lowered or false engagement of handle in the upright or horizontal position.There was 1 event with patient involvement; no adverse consequences were reported.
|
| |
|