• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LGC TIBIAL FIT TRAY CEM SZ 2.5F / 2.5T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LGC TIBIAL FIT TRAY CEM SZ 2.5F / 2.5T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-45-2525
Medical Device Problem Codes Mechanical Problem (1384); Unstable (1667); Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Discomfort (2330); Osteolysis (2377); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 01/17/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 concomitant devices (4617497) 200-02-32 - three peg patella 32mm (4660014) 02-010-01-0225 - logic femoral ps cem left sz 2.5 (4723484) 02-012-45-2525 - lgc tibial fit tray cem sz 2.5f / 2.5t.The product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported via legal documentation that approximately 70 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear significant pain, swelling, depression, difficulty sleeping, scarring, significant weight loss, inability to stand/walk for extended periods or lengths, and reduced mobility.During plaintiff¿s revision surgery, surgeons noted numerous obvious bony defects on the femur and tibia caused by the defective optetrak knee system having been implanted in plaintiff.94.Plaintiff¿s surgeons also observed and noted that plaintiff¿s patella had experienced significant osteolysis (bone loss), to the extent that at the time of plaintiff¿s revision surgery, plaintiff¿s ¿patella was too thin to accept a new patellar component.¿6 the degeneration of plaintiff¿s patella occurred subsequent to the initial tkr wherein defendants¿ optetrak device was implanted.Plaintiff¿s surgeons also noted posterior and distal defects on plaintiff¿s tibia which were not present or observed prior to defendants¿ optetrak knee replacement system being implanted.All of the aforementioned issues observed by plaintiff¿s surgeons during the revision surgery on january 17, 2023 (i.E.Loose components, bone defects, bone loss/osteolysis) were, upon information and belief, directly and proximately caused by premature polyethylene wear and material defects and/or premature failure of the optetrak components manufactured by defendant.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
1038671-2024-04540, 1038671-2024-02587, 1038671-2024-02586 this follow-up report is being submitted to relay additional and/or corrected information.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2024-04540, 1038671-2024-02587, 1038671-2024-02586.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, tibial loosening, femoral loosening, instability, and loss of range of motion, or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGC TIBIAL FIT TRAY CEM SZ 2.5F / 2.5T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville 32653
3523771140
MDR Report Key19860878
Report Number1038671-2024-02588
Device Sequence Number8160325
Product Code JWH
UDI-Device Identifier10885862003935
UDI-Public10885862003935
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K101981
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number02-012-45-2525
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2024
Supplement Date Received by Manufacturer11/19/2024
Initial Report FDA Received Date07/30/2024
Supplement Report FDA Received Date11/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/04/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-