|
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor blueprints found that no burrs, cracks or contamination is allowed for the anchor.Also, a statement (certificate of compliance) or evidence of conformance to material and processing specifications must be provided.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|