BLUE BELT TECHNOLOGIES RI ROBOTIC DRILL ATTACHMENT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB10015
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during cori assisted tka surgery, the ri robotic drill attachment has almost been locked in the drill.The procedure was resumed, after a non-significant delay, changing to manual procedure.No injury was reported as a consequence of this issue.
 
Additional Manufacturer Narrative
Internal reference number: (b)(4).
 
Additional Manufacturer Narrative
The ri robotic drill attachment, part number rob10015, serial number (b)(6), intended for treatment was not returned for evaluation, therefore a device analysis was unable to be performed.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with a loose retaining nut caused by vibration and build up of debris prohibiting the carriage from reaching the unlock position.Based on the investigation, no containment or corrective action is recommended or required at this time.Should any additional information be received, the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.Internal complaint reference number: case (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RI ROBOTIC DRILL ATTACHMENT
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19868862
Report Number3010266064-2024-00163
Device Sequence Number15370235
Product Code OLO
UDI-Device Identifier00885556757345
UDI-Public00885556757345
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberROB10015
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2024
Supplement Date Received by Manufacturer08/14/2024
Initial Report FDA Received Date07/30/2024
Supplement Report FDA Received Date08/21/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORI PN: ROB10024/ SN: (B)(6)
Patient SexUnknown
-
-