| Catalog Number |
0375564000 |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was smoking.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: smoking and turning red.Probable root cause: manufacturing defect, high friction due to components interaction, surface not smooth, poor/lack of suction, clogged shaver blade preventing fluid suction and cooling.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was smoking.
|
| |
|