| Catalog Number |
544965 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the applier jaws were found misaligned before use.Therefore, a new unit was used instead.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The sample was returned to the manufacturer for evaluation.The manufacturer reported: "the dhr for the returned instrument was reviewed and found complete without any irregularities.This instrument was produced at the tecomet, kenosha, wi facility as part of a (b)(4) pc.Lot in june of 2022.It was found that this order was made from the correct materials and components and all approved processes were followed.It can then be stated that the alleged non-conformance inciting this complaint was not due to an error in tecomet -kenosha's manufacturing processes.Evaluation of the returned instrument as received shows that it has a non-teleflex luer port cap installed on it and the jaws are loose and misaligned to each other in the closed position but there is no visible damage to the jaw pivot pin area of the outer tube assembly.After the initial evaluation this instrument was dis-assembled in order to evaluate its internal components and it was found that the inner drive rod (n00185) bosses are both damaged where they engage the jaws.We suspect that the damaged drive rod bosses caused the jaws to become loose and misaligned in the closed position.We are unable to determine what caused the drive rod bosses to become damaged but mishandling such as excessive force being applied to the handle to jaw mechanism of this device at the end user's facility is suspected.All (b)(4) instruments from this lot were 100% visually inspected and function tested and properly lubricated prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed."teleflex will continue to monitor and trend on complaints of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that the applier jaws were found misaligned before use.Therefore, a new unit was used instead.
|
| |
|
Search Alerts/Recalls
|