Catalog Number 544965
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that the applier jaws were found misaligned before use.Therefore, a new unit was used instead.
 
Additional Manufacturer Narrative
(b)(4).The sample was returned to the manufacturer for evaluation.The manufacturer reported: "the dhr for the returned instrument was reviewed and found complete without any irregularities.This instrument was produced at the tecomet, kenosha, wi facility as part of a (b)(4) pc.Lot in june of 2022.It was found that this order was made from the correct materials and components and all approved processes were followed.It can then be stated that the alleged non-conformance inciting this complaint was not due to an error in tecomet -kenosha's manufacturing processes.Evaluation of the returned instrument as received shows that it has a non-teleflex luer port cap installed on it and the jaws are loose and misaligned to each other in the closed position but there is no visible damage to the jaw pivot pin area of the outer tube assembly.After the initial evaluation this instrument was dis-assembled in order to evaluate its internal components and it was found that the inner drive rod (n00185) bosses are both damaged where they engage the jaws.We suspect that the damaged drive rod bosses caused the jaws to become loose and misaligned in the closed position.We are unable to determine what caused the drive rod bosses to become damaged but mishandling such as excessive force being applied to the handle to jaw mechanism of this device at the end user's facility is suspected.All (b)(4) instruments from this lot were 100% visually inspected and function tested and properly lubricated prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed."teleflex will continue to monitor and trend on complaints of this nature.
 
Event or Problem Description
It was reported that the applier jaws were found misaligned before use.Therefore, a new unit was used instead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL ML 5MM ENDO APPLIER
Common Device Name
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
mariah mackinnon
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key19879333
Report Number3011137372-2024-00113
Device Sequence Number14486752
Product Code GDO
UDI-Device Identifier34026704623625
UDI-Public(01)34026704623625(11)220623(17)001231(10)06A2244692(21)
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number544965
Device Lot Number06A2244692
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/31/2024
Supplement Report FDA Received Date08/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-