DAVOL INC., SUB. C.R. BARD, INC. -1213643 HYDRO-SURG PLUS LAPAROSCOPIC IRRIGATOR; LAPAROSCOPE, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Catalog Number |
0026870 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As reported, hydro-surg irrigator male adapter cracked, unable to attach probe tip securely.Evaluation of the sample finds that the probe tip base broke on the qd adapter.The clear portion is fractured diagonally indicating the distal tip was leveraged in one direction forcibly and the clear portion snaped.Based on the sample condition probe tip base broke due to excessive forces applied while in use.Review of manufacturing records confirms product was manufactured to specification, with no indication of a manufacturing related cause for the event reported.This mdr represents one of the hydro-surg plus laparoscopic irrigator used in the case.An additional mdr was submitted to represent another hydro-surg plus laparoscopic irrigator used.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
As reported, during laparoscopic cholecystectomy, the hydro-surg irrigator broke and malfunctioned during a case.It was reported that the two devices, the plastic at the end where the suction tip was cracked (male adapter), making it impossible to get the proper seal and suction.The surgeon tried turning up the suction for the two that cracked but the suction tip would not stay attached.It was also reported that the surgeon used another new device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
As reported, hydro-surg irrigator male adapter cracked, unable to attach probe tip securely.Evaluation of the sample finds that the probe tip base broke on the qd adapter.The clear portion is fractured diagonally indicating the distal tip was leveraged in one direction forcibly and the clear portion snaped.Based on the sample condition probe tip base broke due to excessive forces applied while in use.Review of manufacturing records confirms product was manufactured to specification, with no indication of a manufacturing related cause for the event reported.Addendum: h11: this supplemental mdr is submitted to correct device manufacture date and imdrf annex b code.There is no change to the initial determination.No lot number has been provided; therefore, a review of the manufacturing records is not possible.This supplemental mdr represents one of the hydro-surg plus laparoscopic irrigator used in the case.An additional mdr was submitted to represent another hydro-surg plus laparoscopic irrigator used.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
As reported, during laparoscopic cholecystectomy, the hydro-surg irrigator broke and malfunctioned during a case.It was reported that the two devices, the plastic at the end where the suction tip was cracked (male adapter), making it impossible to get the proper seal and suction.The surgeon tried turning up the suction for the two that cracked but the suction tip would not stay attached.It was also reported that the surgeon used another new device.There was no reported patient injury.
|
| |
|