SYNTHES GMBH SMALL HEXAGONAL SCREWDRIVER SHAFT Back to Search Results
Catalog Number 314.03
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: e3: reporter is a j&j employee.H3, h4, h6: part # 314.03.Synthes lot # a4jm214.Supplier lot # a4jm214.Release to warehouse date: 30 mar 1999.Manufactured by- synthes brandywine.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Photos were provided for review.The photo investigation revealed that small hexagonal screwdriver shaft] ,the provided evidence was not sufficient to confirm the reported event of unable to assemble/disassemble.Functionality issues cannot be evaluated through a photo investigation.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was not confirmed as the observed condition of the small hexagonal screwdriver shaft would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an july 9, 2024, while inventorying the set it was found that the screwdriver shaft were stuck inside the screwdriver handle.There was no patient involvement.This report involves one (1) small hexagonal screwdriver shaft.This is report 1 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device revealed that the small hexagonal screwdriver shaft was found stuck inside the handle with quick coupling, small.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.A functional test was performed where the sleeve slides correctly; however, it does not successfully disengage from the screwdriver.Therefore, the reported condition could be replicated.A dimensional inspection was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the small hexagonal screwdriver shaft would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): part # 314.03, synthes lot # a4jm214, supplier lot # a4jm214, release to warehouse date: 30 mar 1999, manufactured by- synthes brandywine.No ncrs were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL HEXAGONAL SCREWDRIVER SHAFT
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19882811
Report Number8030965-2024-09510
Device Sequence Number17192842
Product Code HXX
UDI-Device Identifier10886982188380
UDI-Public(01)10886982188380(11)990330(17)411231(10)A4JM214
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number314.03
Device Lot NumberA4JM214
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/09/2024
Supplement Date Received by Manufacturer08/19/2024
Initial Report FDA Received Date08/01/2024
Supplement Report FDA Received Date08/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/30/1999
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HANDLE WITH QUICK COUPLING, SMALL
Patient SexUnknown
-
-