Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately 4 years post implantation of a right total hip arthroplasty, the patient is experiencing pain and a loose cup.Subsequently, the patient is being considered for a revision on an unknown date.No additional information was available.
 
Additional Manufacturer Narrative
(b)(4).D10: cat# 00-6250-065-50; lot# 64561604; trilogy bone screw.Cat# 00489405810; lot# 63392067; trabecular metalâ acetabular revision system acetabular augment.Cat# 00662406540; lot# 64645209; bone screw self-tapping 40mm length 6.5 mm dia.Cat# 00-6624-065-50; lot# 63704590; bone scr 6.5x50 self-tap.Cat# 11-300814; lot# 774180; arcos 14x150mm spl tpr dist.Cat# 11-300818; lot# 147420; arcos 18x150mm spl tpr dist.Cat# 010000903; lot# 3743858; g7 10 deg e1 liner 40mm f.Cat# 11-301331; lot# 138850; arcos con sz a hi 70mm.Cat# 650-1058; lot# 2993268; cer bioloxd option hd40mm.Cat# 650-1067; lot# 3031106; cer option type 1 tpr sleve+3.Product remains implanted will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4;d4;g3;h2;h3;h4;h6;h10.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.The reported products were reviewed for compatibility, and it was determined that the following implants are not compatible: the 00625006550 screw with the tmars cup, and the augment, and the g7 liner with the tmars cup.Medical records were not provided.A definitive root cause cannot be determined.This complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2.Upon an internal data review, it was determined that this device was reported under the incorrect manufacturer.This report should be considered void, and the event will be reported under mfr 3005751028.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRABECULAR METAL REVISION SHELL 56MM
Common Device Name
PROTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19885612
Report Number0001822565-2024-02532
Device Sequence Number7283585
Product Code JDI
UDI-Device Identifier00889024310872
UDI-Public(01)00889024310872(17)241130(10)64564812
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K001759
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Model NumberN/A
Device Catalogue Number00700005620
Device Lot Number64564812
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2024
Supplement Date Received by Manufacturer01/14/2025
03/19/2026
Initial Report FDA Received Date08/01/2024
Supplement Report FDA Received Date01/15/2025
03/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-