Catalog Number 199723745S
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Insufficient Information (4580)
Date of Event 07/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: reporter is a j&j employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that an initial surgery (l1/2/3) performed on (b)(6) 2022 with verse fenestrated system.After the surgery, adjacent intervertebral disorders was confirmed.On (b)(6) 2024, the extension surgery to l4 was performed.Before the surgery, there were no fracture reports, no imaging findings, no patient reports, etc.Cement had been injected.In the surgery, the set screw and rod were removed.When the l3 screws in question was pulled out with a core screwdriver to check for loosening of the screw as per the procedure manual, breakage was confirmed around 5 mm from the tip of the screw inserted in l3.The tip was broken by air toming from the extraction site, and a new screw was reinserted and fixed.The broken fragments remain in the vertebral body.The surgery was completed successfully with 30 minutes surgical delay.This (b)(4) is related to (b)(4) which reports an adjacent intervertebral disorders confirmed after the initial surgery.This (b)(4) reports the screw breakage and difficulty of removing the screw occurred in the second surgery.This report is for one (1) 5.5 exp verse fen scr 7.0x45.This is report 2 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9 corrected.H3, h4, h6 the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned device found that the 5.5 exp verse fen scr 7.0x45 was broken from the threaded tip, the broken fragment was not returned for evaluation.Without an x-ray we cannot confirm that a fragment has remained inside the patient.No other issues were observed.A dimensional inspection for the 5.5 exp verse fen scr 7.0x45 was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the 5.5 exp verse fen scr 7.0x45 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a manufacturing record evaluation was performed for the finished device part: 199723745s lot: 345067 it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 25-may-2022 manufacturing site: jabil le locle expiry date:30-apr-2027.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE FEN SCR 7.0X45
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
8472871282
MDR Report Key19899385
Report Number1526439-2024-01991
Device Sequence Number6409434
Product Code NKB
UDI-Device Identifier10705034444505
UDI-Public(01)10705034444505(11)220522(17)270430(10)345067
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200245
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number199723745S
Device Lot Number345067
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/11/2024
Supplement Date Received by Manufacturer08/06/2024
Initial Report FDA Received Date08/02/2024
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.5 EXP VERSE FEN SCR 7.0X45
Patient Age70 YR
Patient SexUnknown
-
-