• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG PRECISION LED LIGHT SOURCE; LIGHT SOURCE, FIBEROPTIC, ROUTINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG PRECISION LED LIGHT SOURCE; LIGHT SOURCE, FIBEROPTIC, ROUTINE Back to Search Results
Catalog Number 0220220000
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was loss of light.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: light output intermittent probable root cause: damaged light cable damaged scope damaged coupler damaged leds main board malfunction loose / misaligned jaw assembly light engine/ light pipe malfunction or damaged bent esst contact inconsistent esst contact camera head communication front board malfunction touch panel malfunction power supply malfunction thermal switch malfunction ac inlet board malfunction inadequate air flow sidne port malfunction poor workmanship inadequate calibration use error electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge the reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was loss of light.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG PRECISION LED LIGHT SOURCE
Common Device Name
LIGHT SOURCE, FIBEROPTIC, ROUTINE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key19899718
Report Number0002936485-2024-00536
Device Sequence Number6556627
Product Code FCW
UDI-Device Identifier07613327051322
UDI-Public07613327051322
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K082813
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0220220000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/09/2024
Supplement Date Received by Manufacturer07/09/2024
Initial Report FDA Received Date08/02/2024
Supplement Report FDA Received Date11/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/01/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-