Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigation summary: the product was returned to ethicon for evaluation.The returned product determined that one detached needle, and one dispensed suture outside its packaging that pertained to product code pdp333h.During the visual inspection of the returned sample, no issues related to breakage sutures were found on the strand.However, short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was detached from the needle since the insertion end of the suture did not meet the requirement.Additionally, a photo was provided and it showed a foil and suture inside a plastic bag.Based on the information currently available, short insertion issue was identified as pull out during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2024 and suture was used.Before use on the patient, it was reported that the suture had been found broken before using on patient during surgery.Changed another one to complete the surgery.During the visual inspection of the returned sample, no issues related to breakage sutures were found on the strand.However, short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was detached from the needle since the insertion end of the suture did not meet the requirement.Additionally, a photo was provided and it showed a foil and suture inside a plastic bag.Based on the information currently available, short insertion issue was identified as pull out during the investigation of the sample received.No adverse patient consequences were reported.Additional information was requested.
|