• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 3000TFX
Medical Device Problem Codes Calcified (1077); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 06/19/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned through implant patient registry and investigation that a 21mm 3000tfx aortic valve was explanted after an implant duration of five (5) years, one (1) month due to calcification and pannus.The patient presented with shortness of breath.The explanted valve was replaced with a 21mm 11500a valve.Patient was taken to the icu post procedure.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated section h6 (type of investigation, investigation findings, investigation conclusions).Calcific degeneration is a common cause of bioprosthetic heart valve failures.Many factors contribute to the onset and propagation of calcification.These include patient factors (age, disease state, pharmacological intervention, etc.), mechanical stress related to the valve's hemodynamic performance, and glutaraldehyde fixation of tissue.Of these, the fixation process is a relatively minor contributor to calcification for edwards' tissue valves due to anti-calcification treatments during manufacturing.Though numerous studies have been conducted on preventive calcification strategies in bioprosthetic heart valves, the causes of calcification are not fully understood and there are still no mechanisms or medical therapies which fully prevent bioprostheses from calcifying.Pannus is a non-immune inflammatory reaction of the body to the implanted prosthesis, characterized by proliferation of fibroelastic tissue and collagen, with a starting point in the suture area and subacute or chronic centripetal evolution.Pannus can have both beneficial and harmful effects depending on the amount of growth and location.A small amount of host tissue growth over the suture line is expected and is needed to form a non-thrombogenic surface and complete the healing process after device implantation.In contrast, an excessive amount of pannus growth can cause nonstructural dysfunction that may require intervention.Host tissue growth can extend onto the cusp surfaces leading to thickening of the cusps, cusp retraction or curling, leaflet immobility, and/or abnormal coaptation, potentially resulting in valvular regurgitation, elevated gradients, and/or stenosis.Calcification is most commonly related to patient factors and is not usually an indication of a device malfunction.The root cause of this event was determined to be due to patient related factors.The event observed in this case was due to a progression of the patient's underlying valvular disease pathology with structural valve deterioration and nonstructural dysfunction, including hyperlipidemia.The subject device is not available for evaluation, as it was discarded.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key19906390
Report Number2015691-2024-05879
Device Sequence Number18325686
Product Code DYE
UDI-Device Identifier00690103163643
UDI-Public(01)00690103163643(17)220903
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/03/2022
Device Model Number3000TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2024
Supplement Date Received by Manufacturer08/27/2024
Initial Report FDA Received Date08/05/2024
Supplement Report FDA Received Date08/30/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/04/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-