|
This report is being supplemented to provide additional information based on the legal manufacturer's (lm) final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported issue was confirmed.However, the definitive root cause could not be determined.Additionally, the results of confirmation of the device and the investigation results in the past, a likely mechanism causing the broken cutting wire might be the following.1.The device was not protruded enough from the endoscope until the rear end of the cutting wire was in the field of view.2.Due to the situation of ¿1¿ description, the cutting wire and the endoscope were being close to each other.3.The output was activated in state of ¿2¿ description.This might have led to an electrical discharge between the cutting wire and the distal end of the endoscope.4.An electrical discharge possibly occurred, and the cutting wire became hot instantly.That might have caused the cutting wire to break.This instruction for use (ifu) manual contains the following information.(drawing no.Rk1134 revision no.03): since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strongly.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the instrument from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.Be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.In case the cutting wire contacts the forceps elevator, insufficient output or unintended tissue injury may occur.Do not activate output when the distal end of the endoscope is too close to or in contact with the cutting wire.This could burn the tissue and could also damage the endoscope and/or instrument.Olympus will continue to monitor field performance for this device.
|