MAUDE Adverse Event Report: KARL STORZ SE & CO. KG ROBI KELLY FORCEPS INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG ROBI KELLY FORCEPS INSERT Back to Search Results
Model Number 38610MD
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Temperature Problem (3022)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a laparoscopic hysterectomy, item sparked inside of patient.There was no one from karl storz to witnessed this.Manager, onsite endoscopy believes it is the insert, and would like all three items evaluated together.Female patient not harmed, case was not delayed.
 
Additional Manufacturer Narrative
The device will be forwarded to the manufacturing site for investigation.Should relevant additional information/investigation results become available, a supplemental medwatch report will be submitted unsolicited.This event is filed under internal complaint id (b)(4).
 
Additional Manufacturer Narrative
The claimed article was visually inspected, and temperature-related damage was found at the distal end.On closer investigation of the article under a photomicroscope it was found that there was a short circuit between the two halves of the jaws, which generated enough heat to melt the plastic in the vicinity of the short circuit.An evaluation was made for this article with the yn02 solder to determine whether several articles with this solder were affected.The result of the evaluation is that there is only one message.Based on this information and the rejected articles it is likely that it is a user error.The high voltage test documented in the test lot was also inconspicuous.Here, both the contact resistance of max.1.0 ohm and the high voltage itself with 700vdc were measured.The result was unobjectionable.It is therefore probable that the ifu was not adhered to in the reprocessing area or that the article was incorrectly applied at the end of the operation.This event is filed under internal complaint id (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROBI KELLY FORCEPS INSERT
Common Device Name
FORCEPS INSERT
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, deu 78523 2
GM  785232
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, deu 78523 2
GM   785232
Manufacturer Contact
anja fair
2151 e. grand avenue
el segundo 
MDR Report Key19912013
Report Number9610617-2024-00238
Device Sequence Number8802993
Product Code GEI
UDI-Device Identifier04048551328682
UDI-Public04048551328682
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K962030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number38610MD
Device Catalogue Number38610MD
Device Lot NumberYN02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2024
Supplement Date Received by Manufacturer10/01/2024
Initial Report FDA Received Date08/05/2024
Supplement Report FDA Received Date10/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OUTER SHEATH ITEM # 38600.; PLASTIC HANDLE ITEM # 38151.
Patient SexFemale
-
-