Model Number 8888345629HP
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: 8888345611hp 12fr 16cm lum hp curv kit (lot#: 2308000099); 8888345611hp 12fr 16cm lum hp curv kit (lot#: 2308000099).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, post-operatively, it appears that there was a weld on the third lumen, which has become broken.The weld part on the third lumen seems to be a fracture point.The remedial action performed was to remove the catheter.There was no luer adapter issue.There was no leak.The catheter was not repaired.The product was not replaced.Nothing unusual is observed on the device prior to use.No other damage was found on the product.There were no other products being utilized with the device.No excessive force use on the device.The insertion site was not treated prior to product placement.Octenisept was the cleaning agent used on the device.There was no blood loss, and blood transfusion was not required.There was no reported patient injury.
 
Additional Manufacturer Narrative
Additional information: h3: evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted one catheter.The luer of the third port had become detached from the tube (and was not pictured).The end of the third port showed evidence of adhesive, and the customer stated that the luer was attached when the device was opened.It was reported that there was a leak or hole on the extension tube at or near the luer adapter.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Common Device Name
CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPL
Manufacturer (Section D)
COVIDIEN LLC
15 hampshire st
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN LLC
15 hampshire st
mansfield MA 02048
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19912416
Report Number1282497-2024-00291
Device Sequence Number8618869
Product Code NIE
UDI-Device Identifier10884521128163
UDI-Public10884521128163
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K102605
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8888345629HP
Device Catalogue Number8888345629HP
Device Lot Number2200400229
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2024
Supplement Date Received by Manufacturer09/04/2024
Initial Report FDA Received Date08/05/2024
Supplement Report FDA Received Date09/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTE ON H11.
Patient Age69 YR
Patient SexMale
Patient Weight89 KG
-
-