| Model Number |
TJF-145 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the duodenovideoscope exhibited foreign material found inside the biopsy channel.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the legal manufacturer's final investigation results.Based on the results of the investigation, a definitive root cause of the foreign material could not be identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The event can be detected/prevented by following the instructions for use (ifu) which state: -ifu states the detection method in instructions evis exera duodenovideoscope olympus tjf type 145 operation manual chapter 3 preparation and inspection.-ifu states the preventative measures in instructions evis exera duodenovideoscope olympus tjf type 145 reprocessing manual chapter 3 cleaning, disinfection and sterilization.Olympus will continue to monitor field performance for this device.
|
| |
|