MAUDE Adverse Event Report: EXACTECH, INC. LOGIC FEMORAL PS CEM RIGHT SZ 3.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC FEMORAL PS CEM RIGHT SZ 3.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-010-01-0335
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 04/16/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(h3) pending evaluation.(d10) concomitant device(s): 6332268 - 02-012-43-3525 - lgc tibia rbktray cem sz 3.5f/ 2.5t.6118219 - 201-78-15 - holding pin mini sharp point 4 pk.5851763 - 201-78-81 - 3 trocar, mod.Hex 2pk.S036859 - 521-78-23 - threaded pin size 2.3 collared.S040394 - 521-78-23 - threaded pin size 2.3 collared.S044527 - 521-78-32 - threaded pin size 3.0 collarless.8011019172 - a10012 - gps implant kit v2.
 
Event or Problem Description
It was reported that this 71 y/o female patient right knee was revised 4 years 2 months post op.Everything was revised due to osteolysis, loosening of femur, patella and tibial insert wear.Patient was last known to be in stable condition following the event.Product not returning - discarded at the hospital.
 
Additional Manufacturer Narrative
(h3) the revision reported was likely the result of presumed prosthesis wear, osteolysis, and an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening.However, images of the revised insert do not show signs of excessive prosthesis wear inconsistent with being implanted.The reported femoral loosening could not be confirmed as the revised devices were not returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM RIGHT SZ 3.5
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19914112
Report Number1038671-2024-02692
Device Sequence Number18226962
Product Code JWH
UDI-Device Identifier10885862001245
UDI-Public10885862001245
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K093360
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/17/2024
Device Catalogue Number02-010-01-0335
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2024
Supplement Date Received by Manufacturer08/07/2024
Initial Report FDA Received Date08/05/2024
Supplement Report FDA Received Date08/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient Weight73 KG
-
-