This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the results of the investigation, the defective event was likely caused by stress of repeated use, external factors or handling of the device; however, a definitive root cause could not be identified.The event can be detected/prevented by following the instructions for use (ifu) which state: 3.2 inspection of the endoscope.5.Inspect the external surface of the entire insertion tube including the bending section and the distal end for dents, bulges, swelling, peeling, scratching, holes, sagging, transformation, bends, adhesion of foreign body, dropout of parts, any protruding objects or other irregularities.Olympus will continue to monitor field performance for this device.
|